Descargar Imprimir esta página

Stryker 1588 AIM Manual Del Usuario página 28

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 290
P P e e r r f f o o r r m m i i n n g g t t h h e e W W h h i i t t e e B B a a l l a a n n c c e e T T e e s s t t
Before each surgical procedure, perform the White Balance test to adjust the camera's perception
of white so it can display other colors correctly.
1. Ensure that a scope and light source are attached to the camera system, and that the console,
light source and monitor are powered on.
2. Point the scope tip at several stacked white gauze pads, a white laparoscopic sponge, or any
clean white surface.
3. Look at the monitor and make sure there is no visible glare off of the white surface of the
image.
4. Press and hold the camera head W button (or "WB" on the touchscreen) until the monitor
displays the message "WHITE BALANCE IN PROGRESS."
5. Continue pointing the scope at the white surface until the video monitor displays the message
"WHITE BALANCE COMPLETE." The image may change color. If you cannot achieve an
acceptable White Balance, refer to the Troubleshooting section.
A A d d v v a a n n c c e e d d F F e e a a t t u u r r e e s s
The 1588 AIM Camera has additional features that are not detailed in this manual:
Button programming
Video image settings
Touchscreen language settings
Light source Run/Standby controls
Other system settings
These advanced features require in-depth knowledge of the device and should be performed only
by trained personnel. For access to advanced features, contact a Stryker representative.
EN
26
P29924F

Publicidad

loading

Este manual también es adecuado para:

1588-010-0001588-210-1051588-610-1221588-710-105