Descargar Imprimir esta página

Stryker 1588 AIM Manual Del Usuario página 32

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 290
P P r r o o b b l l e e m m
SIDNE device does not
recognize camera head
ENV mode not enabled
N N o o t t e e - - If this Troubleshooting section does not resolve the problem, call Stryker Technical
Support at 1-877-478-7953 (inside the U.S.) or refer to the standard warranty.
EN
30
P P o o s s s s i i b b l l e e S S o o l l u u t t i i o o n n
Contact your Stryker representative for compatibility
settings.
Confirm the camera console is set to the
Laparoscopy Surgical Specialty.
Confirm a USB A-to-A cable is connected from the
camera console to the L10 LED Light Source with
AIM Technology
Confirm white light is activated on the light source
P29924F

Publicidad

loading

Este manual también es adecuado para:

1588-010-0001588-210-1051588-610-1221588-710-105