Descargar Imprimir esta página

Stryker 1588 AIM Manual Del Usuario página 31

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 290
P P r r o o b b l l e e m m
Variability in color
reproduction between
different light sources or
peripherals
Foggy picture (loss of
definition and clarity)
Optics are dirty
Blurry picture
Camera head button error
symbol appears on display
monitor:
P29924F
P P o o s s s s i i b b l l e e S S o o l l u u t t i i o o n n
Perform the White Balance test. (See the Performing
the White Balance Test section.)
Check the settings on video peripherals.
Ensure the light source has a proper infrared filter
(check with manufacturer specifications).
Refocus the coupler.
Disassemble the scope, coupler, and camera head,
and clean and dry all windows on the components.
Rotate the scope. If dust particles in the picture
rotate, the dust is located on the scope itself. Follow
the manufacturer's instructions for cleaning the
eyepiece and negative lens.
If particles in the picture do not move when you rotate
the scope, the particles are located on the coupler or
camera. Remove the scope and clean the window on
the front of the coupler with a dry or alcohol-tipped
cotton swab.
Disassemble the scope, coupler, and camera head,
and clean and dry all windows on the components.
Ensure all components are completely dry before
reassembling them, or fogging may result.
Ensure the coupler or C-mount scope is in focus.
On the Home screen, ensure the Surgical Specialty
is not set to FLEXI-SCOPE unless you are using a
flexible scope.
Disassemble the scope, coupler, and camera head,
and clean and dry all windows on the components.
Return the camera head to Stryker for service.
If the symbol appears, the camera head buttons are
non-functional but the camera head will still provide a
video signal to the console. The camera can be
operated from the console.
29
EN

Publicidad

loading

Este manual también es adecuado para:

1588-010-0001588-210-1051588-610-1221588-710-105