To minimize galvanic corrosion, avoid soaking dissimilar metals in close proximity.
•
The device cannot withstand an automated disinfection method.
•
The 1588 camera heads are not autoclavable. Steam sterilizing camera heads that are not
marked A A U U T T O O C C L L A A V V E E will result in product damage.
•
Allow the camera head to cool before connecting it to the console. Connecting the camera
head while it is hot may result in system error.
•
When using Steris® liquid chemical sterilization, remove the camera head from the chamber
once sterilization is complete, or moisture may condense inside the camera head and cause
display defects.
L L i i m m i i t t a a t t i i o o n n s s o o n n R R e e p p r r o o c c e e s s s s i i n n g g
•
Do not cross-sterilize the device. Using multiple sterilization methods may significantly reduce
the performance of the device.
•
Repeated sterilization via Ethlyene Oxide or Sterrad® 100NX® can degrade the product's
cosmetic appearance.
•
Do not leave the device in solutions longer than necessary. This may accelerate normal
product aging.
•
Damage caused by improper processing is not covered by the warranty.
M M a a t t e e r r i i a a l l s s a a n n d d E E q q u u i i p p m m e e n n t t
All materials and equipment required to reprocess the camera head shall be supplied by the user
unless otherwise noted.
I I t t e e m m
A A l l l l p p h h a a s s e e s s
Gloves, eye protection,
etc.
C C l l e e a a n n i i n n g g
Water basin
Enzymatic detergent
Tap water
Syringe
2
Soft-bristle brush
Reverse osmosis/
deionized water
4
P29924F
D D e e s s c c r r i i p p t t i i o o n n
Wear protective equipment as required by the medical
facility and procedure.
Large enough to accommodate camera head without
excessive bending of cable
1
Used in cleaning solution to remove surgical debris
To prepare cleaning solutions
To inject detergent into hard-to-reach areas of device
3
To clean exterior of device or hard-to-reach areas of device
To rinse device
33