Use a syringe to inject 50 mL of the detergent into any crevices and mated surfaces
5 times.
When cleaning the 1588 AIM Camera Head with Integrated Coupler (1588-610-122), brush
between all gaps and crevices while pushing down on the endobody clamp. Continue
brushing all gaps and crevices while releasing the clamp to the initial position.
3. R R i i n n s s e e
Remove the device from the prepared detergent. Rinse the device with reverse osmosis/
deionized water at ambient temperature for 90 seconds or until all visible detergent residue
is removed.
After all visible detergent residue is removed, continue to rinse for 30 seconds.
4. A A u u t t o o m m a a t t e e d d W W a a s s h h
Place the device in the washer on an incline to facilitate drainage.
Program the washer using the following parameters:
P P h h a a s s e e
Pre Wash
Enzyme Wash
Wash 1
Rinse 1
Dry Phase
Filtered pressurized air can be used to aid in drying.
Visually inspect the device for cleanliness, paying close attention to hard-to-reach areas. If
visible soil remains, repeat steps 1– 4.
H H i i g g h h - - L L e e v v e e l l D D i i s s i i n n f f e e c c t t i i o o n n ( ( O O p p t t i i o o n n a a l l ) )
W W A A R R N N I I N N G G - - The device must be sterilized after disinfection. Failure to sterilize the device before
reuse presents an acute infection control risk to the patient.
N N o o t t e e - - For necessary materials and equipment, see the Materials and Equipment table.
The device can be disinfected using a disinfecting solution that has the following active ingredient:
≥ 2.4% glutaraldehyde at 25 °C.
1. Clean and prepare the device as recommended in this user manual. Ensure the soaking cap is
installed.
2. Prepare the disinfecting solution according to the manufacturer instructions.
3. Immerse the device in the solution, filling all mated surfaces and crevices.
4. Soak the device in the solution for 45 minutes.
P29924F
R R e e c c i i r r c c u u l l a a t t i i o o n n
T T i i m m e e
2 minutes
2 minutes
2 minutes
2 minutes
7 minutes
T T e e m m p p e e r r a a t t u u r r e e
Cold water
Hot water
Set point 66 °C
(151 °F)
Hot water
115 °C (239 °F)
D D e e t t e e r r g g e e n n t t T T y y p p e e
N/A
Enzymatic
detergent
Non-enzymatic
detergent
N/A
N/A
37
EN