•
System 1E with S40™ Sterilant
•
System 1 Plus with S40 Sterilant
•
System 1 Express with S40 Sterilant
3. Remove the device from the Steris chamber once sterilization is complete, or moisture may
condense inside the device windows and cause fogging.
4. After sterilization, allow the device to completely dry and cool to room temperature before
reconnecting it to a coupler or the console. Otherwise, the lens can fog during use or the
console can produce a system error.
E E t t h h y y l l e e n n e e O O x x i i d d e e ( ( E E O O ) )
1. Clean and prepare the device as recommended in this user manual. Ensure the soaking cap is
installed.
2. If using a sterilization tray (optional), follow any additional instructions provided with the tray.
Use only trays that are compatible with EO.
3. Double wrap the device (or tray) prior to sterilization.
4. Sterilize the device using the parameters below.
P P r r e e c c o o n n d d i i t t i i o o n n i i n n g g
Wrapping
Temperature
Humidity
Vacuum Set Points
Time
E E x x p p o o s s u u r r e e
Temperature
Humidity
Concentration
Time
A A e e r r a a t t i i o o n n
Temperature
Time
5. After sterilization, allow the device to cool to room temperature before reconnecting it to a
coupler or the console. Otherwise, the lens can fog during use or the console can produce a
system error.
P29924F
Double
55 °C (131 °F)
70% RH
1.3 psia
30 min
55 ± 2 °C (131 ± 5 °F)
70% RH (50–80%) ± 5%
725 mg/L, 100% EO
1 hour
35–39 °C (95–102 °F)
12 hours
39
EN