W W A A R R N N I I N N G G - -
•
Do not use cables or accessories other than those provided with the device, as this may result
in increased electromagnetic emissions or decreased immunity to such emissions.
•
If the device is used adjacent to or stacked with other equipment, observe and verify normal
operation in the configuration in which the device will be used prior to a surgical procedure.
Consult the tables below for guidance in placing the device.
•
Equipment which employs RF communications may affect the normal function of the device.
G G u u i i d d a a n n c c e e a a n n d d M M a a n n u u f f a a c c t t u u r r e e r r ' ' s s D D e e c c l l a a r r a a t t i i o o n n : : E E l l e e c c t t r r o o m m a a g g n n e e t t i i c c E E m m i i s s s s i i o o n n s s
The device is intended for use in the electromagnetic environment specified below. The
customer or the user of the device should ensure that it is used in such an environment.
E E m m i i s s s s i i o o n n s s t t e e s s t t
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC61000-3-2
Voltage fluctuations/
Flicker emissions
IEC61000-3-3
P29924F
C C o o m m p p l l i i a a n n c c e e
Group 1
Class B
Class A
Complies
E E l l e e c c t t r r o o m m a a g g n n e e t t i i c c E E n n v v i i r r o o n n m m e e n n t t - -
G G u u i i d d a a n n c c e e
The device uses RF energy only for
its internal function; therefore, its
RF emissions are very low and are
not likely to cause any interference
in nearby equipment.
The device is suitable for use in all
establishments other than domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic
purposes, provided the following
warning is heeded:
W W a a r r n n i i n n g g : : This system is intended
for use by health care professionals
only. This system may cause radio
interference or may disrupt the
operation of nearby equipment. It
may be necessary to take
mitigation measures, such as
reorienting or relocating the system
or shielding the location.
45
EN