M M a a i i n n t t e e n n a a n n c c e e
I I n n s s p p e e c c t t i i o o n n
Inspect the device on a continual basis for unacceptable deterioration such as (but not limited
to) corrosion, discoloration, pitting, cracked seals, or abnormal noises. If a problem is
observed or suspected, the device should be returned for service.
Inspect all components for cleanliness. If fluid or tissue buildup is present, repeat the above
cleaning and sterilization procedures.
Inspect any cables for cuts and breaks. If visible damage is detected, the device should be
returned for service.
U U s s i i n n g g S S t t e e r r i i l l e e D D r r a a p p e e s s
Using sterile drapes will ensure maximum longevity of the camera. For best results, follow the
instructions provided by the drape manufacturer.
S S t t o o r r a a g g e e
Store the device in a dry, clean, and dust-free environment at room temperatures.
R R e e p p l l a a c c i i n n g g t t h h e e F F u u s s e e s s
W W A A R R N N I I N N G G - - To avoid the risk of fire, use only fuses of the value specified on the fuse label
located on the rear panel of the device.
1. Unplug the power cord from the wall outlet and remove the cord from the device.
2. Unlatch the fuse holder above the AC inlet and remove it. (You may need to press the tab on
the fuse holder with a slender screwdriver to release the latch.)
3. Replace the fuse with the same value and rating.
4. Reinstall the fuse holder until the tab snaps in place.
P P e e r r i i o o d d i i c c M M a a i i n n t t e e n n a a n n c c e e S S c c h h e e d d u u l l e e
W W A A R R N N I I N N G G - -
To ensure safe operation of the device, you should periodically perform the following procedure:
Every 12 months, check the earth leakage current to <500 µA (<300 µA in USA), ground
protective earth impedance to <0.1 ohms, and power consumption less than or equal to rated
power. Use a true RMS digital multimeter and safety analyzer to perform this test.
N N o o t t e e - - Refer calibration and operating difficulties not detailed in this manual to your Stryker
Endoscopy sales representative.