Descargar Imprimir esta página

Garant VG1 Manual De Uso página 21

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 160
Fault
No transillumination
No incident light
Counter shows implausible coordinates
Counter shows no coordinates
Dymo label writer not printing
PC does not boot up or cannot be booted up
10. Cleaning
Disconnect from the power supply before starting cleaning. Clean with a slightly
damp cloth. Do not use chemical, alcohol-based, abrasive or solvent-based cleaners.
11. Storage
After use, cover the unit with the dust cover.
If the unit is to be stored for an extended period, place it in the transport case in a dry
place protected from light and dust at temperatures between +5°C and +40°C and
relative atmospheric humidity between 50% and 70%. Protect stored components
against mechanical shaking and damage.
12. Spare parts
Order original spare parts from the Hoffmann Group Customer Service.
13. Disassembly
B
i
For safe transport, comply with Transport, place of erection [} Page 16].
1. Press the "End measurement software" button (5) to shut down the pre-setting
device VG1.
2. Use the main switch to switch off the pre-setting device VG1.
3. Disconnect the cable connections from the pre-setting device VG1 in the reverse
order described in Connecting the power supply [} Page 16]).
4. Move the tower and anti-rotation arm to their end positions, and fit the transport
restraint for the X-axis and Z-axis. See Remove the transport restraint [} Page 16].
5. Screw four bolts of sufficient length (M16x100) into all corners of the VG1.
6. Per bolt, mount a heavy-duty sling and attach each to one fork of a fork-lift truck.
7. Carefully lift the pre-setting device VG1. Ensure there is one person on the right
and one on the left in order to support the tower and prevent it from tipping.
8. Place the pre-setting device VG1 on the Euro pallet.
9. Remove the four M16x100 screws.
10. Using four M16 screws, screw the pre-setting device VG1 to the pallet with the
help of the supplied brackets.
Apply foam between the granite block and the brackets to prevent scratches and
other damage.
11. Tighten the four M16 bolts with a maximum torque of 5 Nm.
12. Fold in the printer and keyboard holder as far as possible. Apply foam between
the holder and the cover to prevent scratches and other damage.
13. Move the camera arm right to the bottom.
14. Rub lubricant (WD-40) into the tool mounting, remove the taper needle roller
bearing, fill the tool mounting with oiled paper.
15. Wrap the pre-setting device in film, screw the side walls of the case to each other.
16. Fit the cover of the case and screw it into position.
14. Technical data
Designation
Maximum measurement range X
Maximum measurement range Z
Power supply
Weight
Length × width × height
Fine adjustment
Possible cause
LED defective
Cable open circuit
Switched off in the PCTC software
LED defective
Cable open circuit
Switched off in the PCTC software
Movement to reference points failed
Communication problem with the USB measurement sys-
tem box
Adapter incorrectly created or incorrect adapter selected
Communication problem with the USB measurement sys-
tem box
Dymo label writer not switched on
Dymo label writer not connected to the PC
Dymo label writer not installed
No power supply
Loose plug at the PC
Base unit
400 mm
500 mm
100 – 230 V, 50 – 60 Hz
150 kg
1500 mm × 590 mm × 1060 mm
Infinite fine adjustment in the X-axis and
Z-axis
Action
Check status in the software
Check status in the software
Shut down software and reboot in order to repeat the
movement to reference points
Check all plugs at the measurement system box
Check adapter data, or recreate the adapter or recalibrate
the adapter
Shut down software and reboot
Check all plugs at the measurement system box
Check all plug connections
Check whether Dymo label writer is selected in the soft-
ware
Install Dymo label writer
Check all plug connections.
Check whether the power supply plug is inserted at the PC
Designation
Taper needle roller bearing
Radial and axial run-out error
Designation
Interfaces
Operating system
Screen
Display accuracy
Designation
Lens
Resolution
Magnification
15. Disposal
Comply with the national and regional environmental protection and disposal regu-
lations for correct disposal or recycling. Segregate items into metals, non-metals,
composite materials and consumables and dispose of them responsibly. Re-use is
preferable to disposal. Contact Hoffmann Group Customer Services.
Base holder
SK50 with interchangeable SK50 needle
roller bearing sleeve
max. 2 μm
Panel PC with touch screen
1× USB
1× Gigabit Ethernet network port RJ45
socket
WLAN
Windows 10
15" touch screen
0.01 mm
CMOS digital camera
Telecentric
1280 × 800 pixels
17-fold
www.hoffmann-group.com
de
en
bg
da
fi
fr
it
hr
lt
nl
no
pl
pt
ro
sv
sk
sl
es
cs
hu
21

Publicidad

loading

Este manual también es adecuado para:

354760