Configurations; Main Unit; Front View; Side View - Riester RVS-100 Manual Del Usuario

Ocultar thumbs Ver también para RVS-100:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 120
NIBP
Measurement
measurements cannot be taken when the heart rate
is extremely low (less than 40 bpm) or extremely high
(greater than 240 bpm) or if the patient is on a heart-
lung machine. Accurate measurement also cannot be
taken when the following conditions exist:
◊ excessive and continuous patient movement such as
shivering or convulsions;
◊ difficulty detecting a regular arterial pressure pulse;
◊ cardiac arrhythmias;
◊ rapid blood pressure changes;
◊ severe shock or hypothermia that reduces blood flow
to the peripheries;
◊ an edematous extremity.
MRI may lead to vessel damage;
1.3 Configurations
The monitor consists of main unit, NIBP cuff, SpO2 sensor, Tempe-
rature sensor (optional) and printer (optional). It can connect to the
optional RVS-200 Wall Diagnostic Station through DC output. The
connection details are provided the corresponding manual for the
RVS-200 Wall Diagnostic Station.

1.4 Main Unit

1.4.1 Front View

Fig.1-1
1) Physiological alarm visual indicator LED's. When a physiological
alarm occurs, this lamp will light up as defined below:
High level alarm: the lamp quickly flashes red.
Medium level alarm: the lamp slowly flashes yellow.
Low level alarm: the lamp lights yellow without flashing.
2) LCD Touchscreen
3) SpO2 connector
4) NIBP connector
5) USB connector x 2
6) Power button
Press this button to turn on the monitor after AC power is
connected or the battery is installed.
Press and hold for 3 seconds to turn the monitor off.
7) Battery charging indicator LED
On: When the battery is being charged.
Off: When the battery is fully charged or there is no battery
in monitor.
8) Power indicator LED. Status of the LED is specified as follows:
Green: When the AC mains connected.
Orange: When the AC mains not connected and monitor is
powered by battery.
Off: When the AC mains not connected.
9) Well for Temp Probe Cover box (20pcs)
10) Covidien Filac 3000 temp probe

1.4.2 Side View

Right side:
Fig.1-2
Limitations:
Accurate
NIBP
1) Grounding terminal
2) Nurse call connector
3) AC power connector (input)
4) DC power connector (output)
5) USB socket x 2
6) Ethernet LAN Network connector
Caution: Devices connected to this monitor must meet
the requirements of the applicable IEC standards (e.g.
IEC 60950 safety standards for information technology
equipment and IEC 60601-1 safety standards for me-
dical electrical equipment). The system configuration
must meet the requirements of the IEC 60601-1 medical
electrical systems standard. Any personnel who connect
devices to this monitor's signal input/output port is re-
sponsible for providing evidence that the safety certifi-
cation of the devices has been performed in accordance
to the IEC 60601-1. If you have any questions, please
contact Riester. If it is not evident from the equipment
specifications whether a particular device combination
is hazardous--for example, due to summation of leakage
currents—please consult the manufacturers or an expert
in the field to ensure the necessary safety of patients and
proper function of all connected devices.
Left side:
Fig.1-3
1) Integrated Thermal Printer

1.4.3 Rear View

Fig.1-4
1) Speaker

1.4.4 Bottom View

1. Battery compartment
1. Battery compartment
Fig.1-5
Caution: Clean the battery contacts regularly to ensure
optimal electrical contact. Before cleaning, power down
the unit and disconnect it from A/C power. To clean the
contacts, rub with a cotton swab dampened (not dripping
wet) with isopropyl alcohol.
40

Publicidad

Tabla de contenido
loading

Tabla de contenido