Fuse Replacement
Never use a fuse other than the fuse model designated by Ambu. Otherwise,
malfunction or failure of the aBox Duodeno may cause a fire or electric shock hazard.
Be sure to turn the aBox™ Duodeno OFF and unplug the power cable before removing
the fuse from the aBox™ Duodeno. Otherwise, fire or electric shock may result.
If the power fails to come on after replacing the fuses, unplug the power cable
immediately from the AC mains power inlet and then contact Ambu. Otherwise,
electric shock may result.
Do not position the device above the patient. If the front is destroyed due to strong
external impacts, splintering may occur. Falling splinters could cause injury to the patient.
Insert the fuse box into this device until it clicks into position. If the fuse box is
inserted incompletely, the power may fail to come ON or a power failure may occur
during operation.
Care and Storage
After wiping with a piece of moistened gauze, dry the aBox™ Duodeno thoroughly
before using it again. If it is used while still wet, there is the risk of an electric shock.
Do not use the device without thoroughly understanding the IFU. If the system is not
properly prepared before each use, equipment damage, patient and operator injury
and/or fire can occur.
Potential of Fire
The system is not protected against fire and explosion. When using the device in areas
with flammable or explosive gases or in areas with oxygen-enriched air, fire or explosions
may occur in the unit. Do not operate the device in areas with flammable or explosive
gases or gas mixtures. Do not operate the device in an oxygen-enriched environment.
Installation, Repair and Maintenance
The device does not contain any parts that can be repaired by the user. Any
disassembling, change or attempt to repair can lead to injury of patient or user and
damage of the system. Installation, repairs and maintenance must be performed by
Ambu staff or Ambu approved staff. Please read section 12 for more information
regarding troubleshooting.
CAUTIONS
Indicate a potentially hazardous situation which, if not avoided, may result in minor or moder-
ate injury. It may also be used to alert against unsafe practices or potential equipment damage.
General
Do not use a pointed or hard object to press the buttons on the front panel. This may
damage the buttons.
Do not apply excessive force to the aBox™ Duodeno and/or other devices connected.
Otherwise, damage and/or malfunction can occur.
Clean and vacuum dust the ventilation grills of the aBox™ Duodeno using a vacuum
cleaner, when necessary. Keep the ventilation grill clear. Otherwise, the aBox™
Duodeno may break down and get damaged from overheating.
Be sure that this device is not used adjacent to or stacked with other equipment
(other than the components of this system) to avoid electromagnetic interference.
Electromagnetic interference may occur to this device when it is placed near
equipment marked with the following symbol
communications equipment such as cellular phones. If radio interference occurs,
mitigation measures may be necessary, such as reorienting or relocating this device
or shielding the location.
or other portable and mobile RF
7