Care, Maintenance, Transport, Conversion; Care / Re-Use - DJO ARTROMOT-S4 Modo De Empleo

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 172

6. Care, Maintenance, Transport, Conversion

6.1 Care / re-use

The ARTROMOT®-S4 is suitable for re-use if
the following points are observed.
Warning!
m
Shock hazard –
Always remove the power cord from the
wall outlet before cleaning.
Shock hazard, equipment damage –
Liquids must not enter the device or the
programming unit.
• The ARTROMOT®-S4 must be disin-
fected by wiping down with a disinfec-
tant. Thus, it complies with the special
hygiene standards for medical technical
equipment.
• The enclosure and removable arm sup-
port assemblies can be cleaned with
commonly used disinfectants and mild
household detergents.
• Use a damp cloth to wipe the CPM
device down.
Warning!
m
Patient hazard –
Patient contamination
Before using the device on another
patient, be sure to clean and disinfect it
according to the instructions given here.
Note!
The manufacturer recommends using
only special disinfectants approved for
medical devices and with the characteris-
tics specified below under "Caution".
Disinfection with the following disin-
fectants and in compliance with the
instructions for use has been evaluated
and approved by the manufacturer in the
risk assessment:
• DESCOSEPT AF lemon
(Part no: 00-311L-xxx)
BAuA Reg.Nr. N-55153,
Manufacturer: Dr. Schumacher GmbH.
Contact time: 2 minutes minimum
Caution!
m
Equipment damage –
− The plastic material used is not
resistant to mineral acids, formic acid,
phenols, cresols, oxidants and strong
organic or inorganic acids with a pH
value below 4.
− Use only clear disinfectants to prevent
discoloration of the material.
− Do not expose the CPM device to
strong ultraviolet radiation (sunlight)
or fire.
103
-0482

Publicidad

Tabla de contenido
loading

Tabla de contenido