DJO ARTROMOT-S4 Modo De Empleo página 96

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 172
Pauses:
Pauses occur at each programmed maximum value. These are the two pause
points:
Pause up:
Maximum adduction value simultaneous with maximum internal rotation value
and maximum anteversion value
Pause down:
Maximum abduction value simultaneous with maximum external rotation value
and maximum retroversion value
They are adjustable in steps of 1 second between 0 and 30 seconds.
Default setting: no pause
Note!
The pause functions are not available for the Wave motion pattern.
Standby
The standby function is only available in the Isolation motion pattern. With the
standby function, you can determine the parking position of a motor by specify-
ing the percentage of the respective range of motion in a given plane (motor A, B
or C) while movements are performed in the currently active plane.
The stop position at the end of the last cycle (for all three planes of motion) can
be selected in steps of 25% between 0% and 100 % of the programmed maxi-
mum range of motion.
Options are:
Standby Ante / Retro:
While abduction/adduction or internal/external rotation exercises are per-
formed, the stop position for motor A (in the anteversion/retroversion plane of
motion) can be specified as a percentage of the set range of motion for motor A.
Standby Ab-/Adduction:
While ante-/retroversion or internal/external rotation exercises are performed,
the stop position for motor B (in the adduction/abduction plane of motion) can
be specified as a percentage of the set range of motion for motor B.
Standby Internal/External Rotation:
While ante-/retroversion or adduction/abduction exercises are performed, the
stop position for motor C (in the internal/external rotation plane of motion) can
be specified as a percentage of the set range of motion for motor C.
Default: 25% each
Cycles:
The number of cycles defines the number of repetitions of a function.This
parameter is available for motion patterns: isolation, oscillation, stretching.
The number of cycles is adjustable from 5 to 20 repetitions.
Default setting: 5
94

Publicidad

Tabla de contenido
loading

Tabla de contenido