Warning!
m
Shock hazard –
Strictly observe the following warnings.
Failure to do so endangers the lives of
the patient, the user, and other persons
involved.
− Allow the ARTROMOT®-S4 to reach
room temperature before use. If the
device was transported at tempera-
tures below 0°C (32°F), allow it to
stand at room temperature for about
2 hours, until all condensation has
disappeared. If the devices was stored
at temperatures in the maximum
storage temperature range, also allow
it to stand for 2 hours to reach room
temperature.
− The ARTROMOT®-S4 can only be
operated in dry rooms.
− When disconnecting the device from
the power line, remove the plug from
the wall outlet first, before disconnect-
ing the cable from the device.
− When connecting the device to other
equipment or when creating a medical
system, check that the sum of leakage
currents will not cause any hazard.
Please contact DJO if you have ques-
tions in this matter.
− Do not use multiple portable
socket outlets (MPSO) to connect
the device to the power line. The
ARTROMOT®-S4 must be connected
to a properly installed wall outlet
with a non-fused earthed wire.
Before connecting the power cord,
it must be completely unrolled and
placed such that it will not get caught
in the moving parts of the device.
− Before cleaning and service interven-
tions, disconnect the device from the
power line by removing the power
cord from the wall outlet.
− Liquids must not be allowed to enter
the CPM device or the programming
unit. If liquids have entered into the
devices, the ARTROMOT®-S4 must
be immediately checked by a service
technician, before it can be reused.
Warning!
m
Equipment malfunction –
– Magnetic and electrical fields are
capable of interfering with the proper
performance of the device. For this
reason make sure that all external
devices operated in the vicinity of the
device comply with the relevant EMC
requirements. X-ray equipment, MRI
devices, radio systems, cell phones, etc.
are possible sources of interference as
they may emit higher levels of electro-
magnetic radiation.
Keep the device away from such
equipment and verify its performance
before use.
− Refer repair and maintenance to
authorized persons.
Persons are authorized after training
by a specialist trained and commis-
sioned by the manufacturer.
– Route all cables such that they will
not get caught by the moving parts
during operation and that they do not
present a stumbling hazard.
– Inspect the ARTROMOT®-S4 for
damage and loose connections at least
once a year.
Caution!
m
Patient hazard –
Preventing chafing and pressure sores.
When your patient is adipose, very tall
or very short, be sure to prevent chafing
and pressure sores
76