Vaseline) may weaken, damage, or destroy the product and should not be
used at any time.
Make sure that you do not lubricate the HME Holder, HME Cassette or
any accessory which is to be held by the Provox LaryTube. This may lead to
accidental detachment.
2.2 Operating instruction
Insertion (Fig. 3):
Provox LaryTube Standard and Fenestrated:
Gently insert the Provox LaryTube directly into the tracheostoma. Attach these
tubes with either Provox LaryClip or Provox TubeHolder (Fig. 4b-c).
Provox LaryTube with Ring:
Attach a Provox Adhesive and then gently insert the Provox LaryTube through
the adhesive into the tracheostoma (Fig. 4a).
Make sure that the adhesive is centered properly so that the LaryTube does
not press on the sides of the stoma.
Insert the Provox HME System component into the HME Holder.
NOTE: Some patients might experience more coughing after inserting the
LaryTube for the first time or when occluding the HME to speak. This usually
diminishes as the user adjusts to the device. If discomfort persists the LaryTube
should be removed.
Device removal
1.
Hold the Provox LaryTube in place and remove the Provox HME System
component from the tube (Fig. 5).
2.
Remove the Provox LaryTube from the stoma.
NOTE: Insertion and/or removal of the LaryTube may occasionally cause
some slight bleeding, irritation or coughing. Inform your clinician if these
symptoms persist.

2.3 Cleaning and disinfection

CAUTION: DO NOT clean the device while inside the stoma. This can cause
injury. Always remove the device from the stoma before cleaning.
Clean the Provox LaryTube after each use. Do not use any water other than
drinking water to clean and rinse the device.
CAUTION: Be careful not to burn your fingers in the warm cleaning water.
1.
Rinse the Provox LaryTube in drinking water (Fig. 6a).