Do not apply olive oil or products containing petroleum-based lubricants
(e.g. Vaseline®) to the endoscope. These products may cause deterioration of the
endoscope. Use only medical grade lubricants which are free of paraffins and silicone.
•
If remote switches, the suction valve or the insufflation/rinsing valve are not returning
to the OFF position after being pressed, gently pull them upwards to return them to
OFF position.
Endoscopic Accessories
•
Do not use damaged endoscopic accessories. Should any irregularities be
observed during use, take appropriate measures to avoid patient injury and/or
damage to equipment.
1.6. Adverse events
Potential adverse events in relation to Ambu Duodeno System (not exhaustive): Infection/
inflammation (including post-ERCP pancreatitis (PEP), cholangitis, cholecystitis, endocarditis
and sepsis), bleeding, perforation, thermal injuries, stent related adverse events,
cardiopulmonary adverse events, air embolism, anaesthesia related adverse events, nausea,
sore throat, abdominal pain and discomfort.
2. System Description
2.1. System Components
Ambu® aScope™ Duodeno
Sterile and single-use device
The aScope™ Duodeno is not available in all countries. For detailed information,
please contact your local sales office.
2.2. Compatible/Applicable Device
Ambu® aBox™ Duodeno
Reusable device
The aBox™ Duodeno is not available in all countries. For detailed information,
please contact your local sales office.
10
Part number
Distal End Outer
Diameter ø
482001000
41.2 Fr
(13.7 mm)
Part number
485001000US (for US market)
485001000 (for non US markets)
Working channel
Inner Diameter ø
12.6 Fr
(4.2 mm)