Boston Scientific MAESTRO 4000 Manual De Instrucciones página 13

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 38
Remote Rear Panel
This section describes the connections and controls on the rear panel of the Remote shown in Figure 3.
Mains
Power
Inlet
Equipotentiality Terminal
Mains Power Inlet
The Mains Power Inlet must be connected to an AC wall outlet designated as "Hospital Grade" or
"Hospital Only" using a "Hospital Grade" power cord.
Potential Equalization Terminal
• This terminal can be used to verify compliance with 60601-1.
• This terminal is used for connecting dedicated ground leads.
Remote Control Connector
Provides an interface between the Remote and the RF Generator.
Speaker Volume Knob
This knob adjusts the volume of the Remote audio speaker.
Maestro 4000™ Cardiac Ablation System Operator's Manual, Global
VA
POWER INPUT
25 VA
Hz
FREQUENCY
50/60 Hz
AC VOLTAGE
100-120 / 220-240V
Made In USA:
560 Cottonwood Drive
Milpitas, CA 95035
Follow Instructions
For Use
USA
cTUVus Mark indicates compliance to
UL 60601-1 and CAN/CSA 22.2 601.1 M90
covering electrical safety requirements
C
US
for the US and Canada.
871-0189-000 REV A
90818619-01A
Figure 3. Remote Rear Panel
Separate
Collection
Maestro 4000™ Remote Control
RF Cardiac Ablation Remote Control
M00440200
4020
Catalog No.
REF
Date of
YYYY-MM-DD
Manufacture
12345678
Serial Number
SN
Speaker
Volume
Remote
Connector
13
Black (K) ∆E ≤5.0

Publicidad

Tabla de contenido
loading

Tabla de contenido