Description
Pod to RF Generator Cable
Length
Connector
Pod
Dimensions (excluding cable)
• Height
• Width
• Depth
• Weight
Pod Connectors
• Recorder
• Catheter
• Indifferent Electrodes
Pod Recorder Filters
• Low Pass Filters
• Low Frequency Cutoff
High-Current Shutoff
Mechanism
RF Generator Power Output
@ 460 kHz
Impedance
22
Black (K) ∆E ≤5.0
Specification
15 feet (4.6 meters)
14-pin Quick Connector
2.3 in. (5.8 cm)
6.8 in. (17.3 cm)
6.2 in. (15.7 cm)
2.2 lb. (1.0 kg)
Quick Connect Connector
9-pin Quick Connect Connector - Type CF-Defibrillation-Proof
Standard male 2-pin for commercial pads - Type CF-Defibrillation-Proof
Referenced to the Indifferent Electrode
–3 dB at 5 kHz ± 1 kHz
RF output ceases to the patient if more than 1.1 Amp flows in either
Dispersive Pad
• Maximum Output Power:
• 100 W for 100-W Pod
• 150 W for 150-W Pod
• Maximum Output Voltage:
• 132 Vrms (187 Vpk) with 100-W pod
• 162 Vrms (229 Vpk) with 150-W pod
• M aximum Output Current (under normal operation):
1 A per dispersive pad
• M aximum RF power setting based on catheter type used and Control
Mode
• U se catheters that have a rated voltage higher than the RF Generator
maximum output voltage rating.
In STANDBY Mode:
The impedance is measured in the 0-300 Ω range
In DELIVER Mode:
High-power catheters display measured impedance in the 25-300 Ω range
and display "LO" or "HI" outside that range.
Non-high-power catheters display measured impedance in the 50-300 Ω
range and displays "LO" or "HI" outside that range.
Maestro 4000™ Cardiac Ablation System Operator's Manual, Global