Example of a report:
-------------- Pipette Calibration --------------
User
Customer
Pipette
Serial no.
Channels quantity
Channel no.
Measurements quantity
Operation with adherence
to ISO 8655
Start date
2015.11.15 07:50:44
End date
2015.11.15 07:54:34
Water temperature
Temperature
Humidity
Pressure
Z coefficient
----------- Tested Volume: 1000
1 0.998 g
2 0.998 g
3 0.998 g
4 0.998 g
5 0.998 g
6 0.998 g
7 0.998 g
8 0.998 g
9 0.998 g
10 0.998 g
Average volume
Average [%]
Systematic error
Systematic error [5]
Permissible error
Random error
Random error [%]
Permissible error
Status
------------------------------------------------------
Signature
Smith
Williams
p901\1k
7777
1
1
10
Yes
o
22.15
C
o
21
C
48%
1005 hPa
1.00328
---------
µl
1000.82389 µl
1000.82389 µl
1000.82389 µl
1000.82389 µl
1000.82389 µl
1000.82389 µl
1000.82389 µl
1000.82389 µl
1000.82389 µl
1000.82389 µl
1000.82389 µl
100.08 %
0.82389 µl
0.08239 %
± 16 µl
0 µl
0 %
± 6 µl
Positive
- 107 -