It is possible to control ionizer using quick access button.
Press to swap between activate and deactivate state.
Press to activate.
Press to deactivate.
The operation is signalled by blue LED diode (installed at the back of the weighing chamber) which
emits light when the ionizer is on.
4.
START-UP
1. Connect the power supply to the mains, ON/LOAD
panel lights up.
2. Press
key, the system and RADWAG program are loaded which is signaled with blinking
of red ON/LOAD diode.
3. Upon PLUS balance start-up the home screen is displayed, levelling process runs automatically,
during
the
automatic
Respective messages are displayed.
Levelling range is limited. In case of too great tilt of the surface (counter, table top) where the
balance rests, the levelling gets aborted and the following message is displayed.
Shall the surface tilt be too great, first level the surface, next carry out the balance levelling (for
levelling procedure details read further down this manual).
Upon completed start-up of the remaining balance series, the home screen is displayed
automatically.
4. The balance runs with no user logged in. In order to start operation it is necessary to log in (for
detailed logging procedure read later sections of this user manual).
Caution: Before you start the balance, make sure that there is no load on the weighing pan.
In accordance with the EN 45501 standard verified balances cannot display mass value below -20e.
If
the
indication
Press
key to zero the balance.
levelling
value
is
below
the
feet
engines
-20e,
<Lo
mass>
- 35 -
diode on the operation
operation
is
information
is
displayed.
heard.