Response format:
1
2
3
U
H
space
Mass
- 9 characters, right justification
Unit
- 3 characters, left justification
Set Mass Value of a Single Item (for PARTS COUNTING mode exclusively)
Format:
SM_XXXXX CR LF, where: _ - space, XXXXX - mass format
Response options:
SM_OK CR LF
- command carried out
- command understood but not accessible at this moment (e.g. other than PARTS
SM_I CR LF
COUNTING mode operated)
ES CR LF
- command not recognised (mass format incorrect)
Set Target Mass Value (e.g. for DOSING mode)
Format:
TV_XXXXX CR LF, where: _ - space, XXXXX - mass format
Response options:
TV_OK CR LF
- command carried out
- command understood but not accessible at this moment (e.g. WEIGHING mode
TV_I CR LF
operated)
ES CR LF
- command not recognised (mass format incorrect)
Set Reference Mass Value (e.g. for PERCENT WEIGHING mode)
Format:
RM_XXXXX CR LF, where: _ - space, XXXXX - mass format
Response options:
RM_OK CR LF
- command carried out
- command understood but not accessible at this moment (e.g. other than
RM_I CR LF
PERCENT WEIGHING mode operated)
ES CR LF
- command not recognised (mass format incorrect)
Value Release
Format: SS CR LF
Response options:
SS_OK CR LF
- command understood and in progress
Command's function is similar to function of PRINT key located on the balance panel.
Internal Adjustment
Format: IC CR LF
Response options:
IC_A CR LF
- command understood and in progress
IC_D CR LF
- adjustment completed
IC_A CR LF
- command understood and in progress
IC_E CR LF
- time limit exceeded while waiting for a stable measurement result
IC_I CR LF
- command understood but not accessible at this moment
Disable Automatic Internal Adjustment of the Balance
Format: IC1 CR LF
Response options:
IC1_I CR LF
- command understood but not accessible at this moment
IC1_E CR LF
- operation disabled, e.g. for verified balance
4-12
13
mass
space
- 189 -
14
15
16
unit
17
18
19
space
CR
LF