1. Enter <
/
2. Enter <
Products> database and press respective record.
List of parameters defined for control:
Pictogram
Data
Name
Code
EAN code
Mass
Tare
PGC mode
Batch portion
PGC: Unit
Batch quantity
Interval
average
determination
Packaging
quantity
Average
coefficient
Internal control
Density
Apart from the above listed options there are other ones which may be used for different working
modes (e.g. CHECKWEIGHING, THRESHOLDS, etc.)
List of parameters defined for internal criteria:
Internal control
Sample Quantity
[T1-] error value
[T1+] error value
Disqualifying quantity of
samples [Qn – 2T]
Databases> submenu.
Description
Product name
Product code
Product EAN code
Product nominal mass
Product tare (set automatically upon selecting a product from the
database).
Control type: Non-destructive average tare, Non-destructive
empty – full, Destructive full – empty, Destructive empty –full
Measurement series for control: Non-destructive empty – full,
Destructive full – empty, Destructive empty – full
Unit of a product: [g] or [ml].
Declaring quantity of controlled batch
of
Time interval given in [h] determining how often average tare
tare
must be estimated for a given product.
Declaring quantity of packages that are to be subjected to
process of average tare determination (for Non-destructive
average tare type of control)
tare
Coefficient for threshold of acceptance of standard deviation for
average tare. Coefficient range: 0.10 - 0.25.
Submenu for determining internal criteria for a control (see: table
below).
Product density (value range: 0.1g/cm
<YES> - enable internal control criteria, <NO> - disable internal control
criteria.
Sample quantity for a product
Max permissible value of –T error given in unit set for a product.
Measurements lower than Qn-T will be considered incorrect.
Max permissible value of +T error given in unit set for a product.
Measurements higher than Qn+T will be considered incorrect.
Quantity of -2T errors disqualifying tested sample.
- 124 -
3
3
- 5g/cm
).