requirements of the applied quality system. The obtained values are entered to balance software:
<Minimum sample weights> database.
For a particular method of minimum sample weight determination, it is possible to define few tare
values, along with values of minimum sample weights, expiry dates of the carried out measurements
and entered data. The settings are not to be edited by a balance user.
<Minimum sample weight> function guarantees that the results of weighing operation are comprised
within the set tolerance, in accordance with applied quality management system of particular
company.
Caution! Function valid for weighing mode exclusively.
Available options:
•
METHOD
Option allowing you to designate used quality standard. Press METHOD field to open a
window with a list of database-stored methods by means of which the minimum sample
weights were determined. In order to add a new method go to <Databases /Minimum
sample weights>.
•
MODE
Block – select to display pictograms informing you on status of weighed mass (weighed
mass below/above minimum sample weight) and to disable accepting the measurement
result if it is below the minimum sample weight limit.
Warn – select to enable display of respective pictograms informing you about the measured
mass (whether it is lower than the MSW value or higher than the MSW value). You can
accept the measurement that is lower than the MSW value. Mind that on a printout it will be
preceded by an asterisk (*).
Caution! For information on how to enter a method for minimum sample weight go to section 29.14
<Databases /Minimum sample weights> of this user manual. New minimum sample weight values can be
entered by an authorized RADWAG personnel exclusively.
Weighing with use of <MINIMUM SAMPLE WEIGHT> function
If in the course of weighing you want to find out whether a particular measurement value is higher
than the MSW value, specified for a given weighing range, then <Minimum sample weight> function
must be on. To turn the function on go to weighing mode settings.
Procedure:
1. Press grey workspace area.
2. The following submenus are displayed: Settings, Buttons, Information, Printouts, Profile.
3. Select <Settings> menu - additional weighing process options.
4. Press < Minimum sample weight> field.
5. Wait for Minimum sample weight window to open, press <Method> field,
window with a list of database-stored methods opens.
6. Select respective method.
7. Previous menu is displayed.
8. Press <Mode> parameter.
9. Settings
Block – select to display pictograms informing you on status of weighed mass (weighed
mass below/above minimum sample weight) and to disable accepting the measurement
result if it is below the minimum sample weight limit.
Warn – select to enable display of respective pictograms informing you about the measured
mass (whether it is lower than the MSW value or higher than the MSW value). You can
box
is
displayed,
select
- 67 -
respective
option: