Switch off Continuous Transmission in Basic Measuring Unit
Format: C0 CR LF
Response options:
C0_I CR LF
- command understood but not accessible at this moment
C0_A CR LF
- command understood and carried out
Switch on Continuous Transmission in Current Measuring Unit
Format: CU1 CR LF
Response options:
CU1_I CR LF
- command understood but not accessible at this moment
CU1_A CR LF
- command understood and in progress
MASS FRAME
- response: mass value in current measuring unit
Response format:
1
2
3
stability
S
U
I
marker
Switch off Continuous Transmission in Current Measuring Unit
Format: CU0 CR LF
Response options:
CU0_I CR LF
- command understood but not accessible at this moment
CU0_A CR LF
- command understood and carried out
Set Min Checkweighing Threshold
Format:
DH_XXXXX CR LF, where: _ - space, XXXXX - mass format
Response options:
DH_OK CR LF
- command carried out
ES CR LF
- command not recognised (mass format incorrect)
Set Max Checkweighing Threshold
Format:
UH_XXXXX CR LF, where: _ - space, XXXXX - mass format
Response options:
UH_OK CR LF
- command carried out
ES CR LF
- command not recognised (mass format incorrect)
Give value of min checkweighing threshold
Format:
ODH CR LF
Response: DH_MASS CR LF - command carried out
Response format:
1
2
3
D
H
space
Mass
- 9 characters, right justification
Unit
- 3 characters, left justification
Give Value of Max Checkweighing Threshold
Format:
OUH CR LF
Response: UH_MASS CR LF - command carried out
4
5
space
character
4-12
13
mass
space
6
7-15
16
mass
space
14
15
unit
- 188 -
17
18
19
unit
16
17
18
space
CR
20
21
CR
LF
19
LF