THRESHOLD:
Mass value obligatory for operation of automatic printout, set in grams.
INTERVAL
Record frequency for automatic operation with interval
Printout
Printout type related to a working mode. Printout is activated by pressing PRINT key on
balance panel.
Available options:
STANDARD PRINTOUT:
Enables declaring content of the following printouts: HEADER, GLP PRINTOUT and
FOOTER.
Select <YES> parameter for the components that are to be printed on pressing
printout activating function key.
NON-STANDARD PRINTOUT:
Enables selecting one of database-stored non-standard printouts form a list of
<PRINTOUTS> menu, or designing a unique printout which is automatically added
to the database.
Caution! For detailed instruction refer to section 14.7 of this user manual.
Air buoyancy compensation
Parameters allowing you to switch on the compensation, and to enter weighed sample density
data and air density data.
Caution! Function valid for weighing mode exclusively. For detailed information regarding the
operation and settings refer to the previous section of this user manual.

15.8. Minimum Sample Weight

Weighing mode settings comprise <Minimum sample weight> function. In order to use the function it
is necessary to enter data related to minimum sample weight determination methods, and minimum
sample weight values for a given method. The data must be entered to <Minimum sample weight>
database. The database is empty in standard balance version.
Determining minimum sample weight and adding data to the <Minimum sample weights>
database can be carried out by an authorized "RADWAG" personnel exclusively.
Should you use the minimum sample weight function, and need data on minimal sample weight to be
entered to balance menu, please contact the local representative of "RADWAG" company.
Using reference weights, the authorized RADWAG employee determines minimum sample weight for
specified tare containers. This operation is carried out at the place of use and in accordance with
– first stable measurement is recorded upon loading of the weighing
pan with a load; record of the next stable measurement result takes
place only if the following conditions are met: weighing pan gets
unloaded, measurement value lower than value of set threshold is
displayed, a new load is placed on the weighing pan.
– last stable measurement, registered prior unloading, is recorded.
The record takes place only if the following conditions are met:
weighing pan gets unloaded, measurement value lower than value of
set threshold is displayed.
– select this value to start automatic printout and record of
indications in WEIGHINGS database, carried out in a cyclic manner
in a specified time intervals. The interval is set in [min] in
<INTERVAL> parameter. Interval range is 1 s – 9 h 59 min 59 s.
Caution!
Each weighing result is printed and recorded (stable and unstable for a
non-verified balance, stable for a verified balance).
Automatic operation with interval starts at the moment of switching the
function on. The first stable weighing result of value greater than
THRESHOLD value is printed and recorded as the first measurement.
The following measurements are printed with frequency set in
INTERVAL parameter. To stop the automatic operation with interval,
switch the option off.