Send Stable Measurement Result in Current Measuring Unit
Format: SU CR LF
Response options:
SU_A CR LF
- command understood and in progress
SU_E CR LF
- time limit exceeded while waiting for stable measurement result
SU_I CR LF
- command understood but not accessible at this moment
MASS FRAME
- response: mass value in basic measuring unit
Response format:
1
2
3
S
U
space
Example:
S U CR LF
S U _ A CR LF
S U _ _ _ - _ _ 1 7 2 . 1 3 5 _ N _ _ CR LF
where: _ - space
Immediately Send Measurement Result in Current Measuring Unit
Format: SUI CR LF
Response options:
SUI_I CR LF
- command understood but not accessible at this moment
MASS FRAME
- command carried out, immediate response: mass value in basic measuring unit
Response format:
1
2
3
stability
S
U
I
marker
Example:
S U I CR LF
- command sent from a computer
S U I ? _ - _ _ _ 5 8 . 2 3 7 _ k g _ CR LF
where: _ - space
Switch on Continuous Transmission in Basic Measuring Unit
Format: C1 CR LF
Response options:
C1_I CR LF
- command understood but not accessible at this moment
C1_A CR LF
- command understood and in progress
MASS FRAME
- response: mass value in basic measuring unit
Response format:
1
2
3
S
I
space
4
5
stability
space
character
marker
- command sent from a computer
- command understood and in progress
4
5
space
character
4
5
stability
space
character
marker
6
7-15
16
mass
space
- command carried out, response: mass value in
current measuring unit.
6
7-15
16
mass
space
- command carried out, response: mass value in
basic measuring unit
6
7-15
16
mass
space
- 187 -
17
18
19
20
unit
CR
17
18
19
20
unit
CR
17
18
19
20
unit
CR
21
LF
21
LF
21
LF