UT_OK CR LF
- command carried out
UT_I CR LF
- command understood but not accessible at this moment
ES CR LF
- command not recognised (tare format incorrect)
Caution:
Use dot in tare format as decimal point.
Send stable measurement result in basic measuring unit
Format: S CR LF
Response options:
S_A CR LF
- command understood and in progress
S_E CR LF
- time limit exceeded while waiting for stable measurement result
S_I CR LF
- command understood but not accessible at this moment
MASS FRAME
- response: mass value in basic measuring unit
Response format:
1
2-3
stability
S
space
marker
Example:
S CR LF
- command sent from a computer
S _ A CR LF
- command understood and in progress
S _ _ _ _ - _ _ _ _ _ _ 8 . 5 _ g _ _ CR
where: _ - space
Immediately Send Measurement Result in Basic Measuring Unit
Format: SI CR LF
Response options:
SI_I CR LF
- command understood but not accessible at this moment
MASS FRAME
- command carried out, immediate response: mass value in basic measuring unit
Response format:
1
2
3
S
I
space
Example:
S I CR LF
- command sent from a computer
S I _ ? _ _ _ _ _ _ _ 1 8 . 5 _ k g _ CR LF
where: _ - space
4
5
space
character
LF
4
5
stability
space
marker
6
7-15
mass
space
- command carried out, response: mass value in
basic measuring unit
6
7-15
character
mass
space
- command carried out, immediate response: mass
value in basic measuring unit
- 186 -
16
17
18
19
unit
16
17
18
19
unit
20
21
CR
LF
20
21
CR
LF