4. In balance with automatically opened doors, door opening and closing test is carried
out.
Autotest GLP function enables:
Report display and printout,
Record of a report containing basic balance data, ambient conditions values and
test results.
Test results:
*Deviation for maximum capacity.
*Value of repeatability expressed as standard deviation.
*Door test result (positive / negative).
Report example:
.................................................................
----------- Autotest GLP: Report ------------
Balance type
Balance S/N
Operator
Software version
Date
Time
------------------------------------------------------
Measurements quantity
Reading unit
Internal weight mass
Filter
Value release
------------------------------------------------------
Deviation for Max.
Repeatability
Signature
.............................................................
Procedure:
1. Press entry with test name, a message box is displayed.
2. Do one of the following:
Start subsequent AUTOTEST GLP process,
Preview results of completed autotests, and export all saved test results as file with *.csv
extension,
Delete all saved test results.
You can abort the process at any moment by pressing <X> button.
Results of carried out autotests are displayed in a table (each line features date and time of autotest
procedure and its result).
To preview particular autotest data, press a respective table line.
In order to print a result of a single autotest, enter its details (see above) and generate a printout by
pressing <Print> key located on the top of the display.
Autotest results can be exported by pressing <Export> entry in the window with all saved reports.
Data is sent to a USB flash drive, connected to balance USB port, as a file with *.csv extension.
XA 4Y
400010
Admin
L0.0.21 S
2012.01.16
09:17:16
10
0.0001 g
140.094 g
Average
Fast and precise
-0.0118 g
0.00088 g
- 175 -