13.2.1.
Settings
Name
Upon entering this parameter, a message box with an on-screen keyboard is
displayed. Enter the name of a profile and press
name is assigned to the profile.
Default working mode
Upon entering this parameter you can select a specific working mode that is to be
set as default. For <None> value set, upon selecting the profile, the balance
activates the most recently used working mode.
13.2.2.
Working Modes
Upon entering Working Modes parameter, a window containing all accessible working modes is
displayed. Parameters are set for each working mode independently. Particular profile when
activated runs with given set of parameters.
List of parameters common for each working mode:
•
Settings: specific parameters relating to a working mode, and universal settings such as:
result control, tare mode, automatic footer printout, printout mode, printout.
•
Functions of quick access buttons: declaring quick access buttons to be displayed.
•
Information: declaring information to be displayed.
•
Printouts: declaring type of a printout or defining a non-standard printout.
13.2.3.
Readout
You can adjust the balance to ambient conditions (filter level) or to your own needs.
List of <Readout> menu parameters:
Filter (disabled for Fast, Fast dosing, Precision profiles)
Each measurement signal before being displayed is electronically processed in order to
provide correct parameters specific for stable measurement result ( ready to be read).
You can determine signal processing rate, there are five different filters to select from:
•
very fast,
•
fast,
•
average,
•
slow,
•
very slow.
While setting the filter consider the actual operating conditions. In case of very good
operating conditions select average or fast filter, in case of rough conditions select slow
or very slow filter.
Caution! For precision balances it is recommended to select very fast, fast or average filter.
For analytical balances and microbalances it is recommended to select average, slow, or
very slow filter.
- 52 -
button to confirm. The