Manual Printout / Automatic Printout - RADWAG 4Y Serie Manual Del Usuario

Ultaramicrobalanza. microbalanza. microbalanzas para pipetas. microbalanzas para los filtros. balanzas analíticas. balanzas analíticas para los filtros
Ocultar thumbs Ver también para 4Y Serie:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Hidden
digits
- Number specifying quantity of hidden digits: [space] when no digits are
quantity
Example:
NT CR LF
- command sent from a computer
N T _ ? _ _ 0 _ _ _ _ _ - 5 . 1 1 3 _ g _ _ _ _ _ _ _ 0 . 0 0 0 _ g _ _ _0 CR
- command carried out, response: mass value and remaining data
where: _ - space

39.3. Manual Printout / Automatic Printout

It is possible to generate printouts either manually or automatically.
Manual printout: wait for a stable indication and press
Automatic printout is generated automatically in accordance with parameters set for automatic
printout (read section 14.7).
The content of a printout depends on settings of <Standard printout> parameter - <Weighing printout
template> (read section 14.7).
Mass printout format:
1
2
stability
space
marker
Stability marker
[space] if stable
[?] if unstable
[!] if air buoyancy compensation function is on
[^] if high limit is out of range
[v] if low limit is out of range
Character
[space] for positive values
[-] for negative values
Mass
9 characters with decimal point, right justification
Unit
3 characters, left justification
Example:
_ _ _ _ _ _ 1 8 3 2 . 0 _ g _ _ CR LF - a printout generated, with reference to <Weighing printout
template> settings, from a balance upon pressing
N
quantity)
Date
Time
Level status
Customer
Warehouse
Product
Packaging
hidden, [1] when 1 digit is hidden
3
4 -12
character
mass
(measurements
NO
NO
NO
NO
NO
NO
NO
NO
13
14
15
space
unit
key:
Universal variable 1 ... 5
Net
Tare
Gross
Current result
Supplementary unit
Mass
Non-standard printout
- 199 -
LF
button.
16
17
18
CR
LF
NO
NO
NO
NO
NO
NO
YES
NO

Publicidad

Tabla de contenido