u1, u2, next
Caution:
If x=next, the command swaps to adjacent unit available on the list (it simulates either 'unit' button
pressing or pressing of unit field displayed in mass window).
Example:
Command:
US_mg<CR><LF>
Response:
US_mg_OK<CR><LF>
Give Current Unit
Command overview:
Command returns current unit.
Format: UG <CR><LF>
Response options:
UG_x_OK<CR><LF>
UG_I <CR><LF>
x - parameter, units symbols
Example:
Command:
UG<CR><LF>
Response:
UG_ct_OK<CR><LF>
Cooperation with PUE 7.1, PUE 10 Indicators
Format: NT CR LF
Response options:
ES CR LF
- command not recognised (mass format incorrect)
MASS FRAME
- response: mass value in basic measuring unit
Response format:
NT
- command
Stability marker
- [space] if stable, [?] if unstable
Zero marker
- [space] for any value but zero, [Z] for zero value
Range marker
- mark of range within which mass value is comprised: [space] I-range balance,
Last digit marker - [zero] no digit, [1] one digit, [2] two digits, [3] three digits, [4] four digits, [5]
Mass
- 10 characters for net mass given in adjustment unit (with dot and a floating "-"
Mass unit
- 3 characters, left justification
Tare
- 9 characters with dot - right justification (when floating tare gets turned off
Tare unit
- 3 characters, left justification
- command carried out, response: currently set unit
- command understood but not accessible at this moment
[2] II-range balance, [3] III-range balance
five digits
sign, no last digit markers), right justification
automatically then zero value is sent)
- set „mg" unit
- „mg" set as a current unit
- return current unit
- currently set unit is „ct"
- 198 -