XA 4Y balance with a set for pipettes calibration:
Set A
Set B
Set C
Installation procedure:
1. disassemble the standard weighing
pan and the anti-draft shield,
2. place centering ring (8) inside the
chamber,
3. put glass ring (7) onto the centering
ring,
4. place weighing pan (6) inside the
glass ring,
5. inside,
vessel (5), put it onto the weighing
pan,
6. if the vessel, used in set C, is too
light, it is necessary to apply load
ring (9) or (10) onto the weighing
pan,
7. put top ring (4) onto the glass ring,
8. put evaporation ring (3) onto the top
ring,
9. put glass lid (2) onto the top ring,
10. put glass lid (1) onto the glass lid.
The XA 4Y.A.PLUS balance with pipette
calibration set installed.
The XA 4Y.A. balance with pipette
calibration set installed.
- 99 -
place
pipette
calibration