accept the measurement that is lower than the MSW value. Mind that on a printout it will be
preceded by an asterisk (*).
10. Set the parameters and exit the menu.
11. Section presenting the weighing result comprises additional pictogram providing
supplementary information. The pictogram changes in the course of weighing operation, it
informs you on the weighed sample mass; the information is referred to the declared
minimum sample weight value.
Weight value lower than the MSW value
specified for a particular tare range.
MSW pictograms:
Weight value lower than the specified MSW value.
Weight value higher than or equal to the specified MSW value.
Weight value lower than the specified MSW value; a pictogram with a clock informs on
expiry of the minimum sample weight validity period (it is displayed two weeks before
the determined expiry date).
Weight value higher than the specified MSW value; a pictogram with a clock informs on
expiry of the minimum sample weight validity period (it is displayed two weeks before
the determined expiry date).
The validity period of the minimum sample weight has expired. The minimum sample
weight settings need to be modified. The minimum sample weight settings can be
modified by an authorized RADWAG personnel exclusively.
Caution! If more than one reference tare value is programmed (with corresponding minimum sample
weight mass) then the indicated value automatically moves to a range corresponding to the weight of the
tared container. At the same time the required minimum mass is changed.
Weight value higher than the MSW value
specified for a particular tare range.
- 68 -