5. Press <START
displayed.
Caution!
If prior to control:
•
Load has not been removed from the weighing pan or if other zeroing requirements have not
been met (e.g. unstable indication), balance shows message: <-Err 2>. Upon noticing <-Err 2>
message unload the balance and wait until all zeroing requirements are met.
•
You have not logged in or you are logged but not assigned with the right permissions to perform
control, message: <Access denied> is displayed.
•
You have not selected product from the database, message: <Product is not selected> is
displayed.
28.6. Abort Control Procedure
You can abort control at any moment.
Procedure:
1. Press
button (abort control) located in the bottom part of the process window.
2. The following message is displayed:
3. Press
button to return to the ongoing control. Press
return to PGC mode.
Control reports with status <Aborted> is saved to PGC database.
28.7. Aborted Control Restoring Procedure, Power Loss Case
In case of disconnection from the mains, to prevent loss of data for control being in progress, the
program continuously records control results.
For power loss case, the software allows to complete aborted control upon reconnection to the
mains.
Caution! Remember to restart the balance when unloaded. It is obligatory to run the balance with an
empty weighing pan.
During balance restart the software searches for any in course control procedure. If there is one, the
following message is displayed:
> button located in the bottom. The home screen containing entered data is
- 126 -
button to abort control and