30.14.
Minimum Sample Weight
Minimum sample weight database contains data concerning declared methods and minimum sample
weights.
Caution! Only authorized RADWAG personnel can add new minimum sample weight records or make
changes in those stored in the database.
Procedure:
1. Enter <
Databases> submenu and press <
2. Press <
Add> button to add a new minimum sample weight, or
3. Select already existing weight.
List of parameters defined for minimum sample weights:
1.
Name
2.
Code
3.
Description
4.
Subsequent control
5.
Thresholds
Example no. 1, XA 220.4Y balance with d=0.0001 g:
List of MSW values determined for the following packaging values:
No.
Tare value
1
0.0000 g
2
10.0000 g
3
50.0000 g
4
200.0000 g
[name of a method used to determine minimum sample
weight]
[method code]
[method overview]
[a field for entering expiry date of the minimum sample
weight. 2 weeks before the expiry date, a pictogram of a
clock is displayed next to the status pictogram; the clock
pictogram informs on elapsing validity date; contact
RADWAG representative to carry out necessary changes in
minimum sample weight settings]
[parameter for entering minimum sample weight data and
range of packaging mass (tare) with applicable criteria:
Tare – maximum tare value, for which the minimum sample
weight
You can enter three characteristic quantities: 0.000g,
optional mass value from weighing instrument's measuring
range, and maximum capacity (see examples below).
Minimum mass – value of the minimum sample weight
determined for a particular weighing device on site using
respective methods.]
Minimum sample
Overview
weight
0.1000 g
Minimum sample weight refers to all net weight values
when weighing is carried out without use of a tare
container (<TARE> button NOT used).
1.0000 g
Minimum sample weight refers to all net weight values
when weighing is carried out using a tare container of
mass ranging from 0.0001 g to 10.0000 g inclusive
(<TARE> button used).
2.5000 g
Minimum sample weight refers to all net weight values
when weighing is carried out using a tare container of
mass ranging from 10.0001 g to 50.0000 g inclusive
(<TARE> button used).
4.0000 g
Minimum sample weight refers to all net weight values
when weighing is carried out using a tare container of
mass ranging from 50.0001 g to 200.0000 g inclusive
(<TARE> button used).
Minimum samples weights> entry.
is
- 151 -
obligatory.