22.1. Mode Related Settings
The supplementary settings enable you to adjust the working mode to your needs and requirements.
To access the settings follow the below procedure.
Procedure:
1. Press grey workspace area.
2. The following submenus are displayed: Settings, Buttons, Information, Printouts, Profile.
3. Select <Settings> menu.
4. Functions related to formulations are displayed.
Formulation mode settings:
•
Automatic ingredients naming:
•
Yes
•
No
•
Apply tare from database:
•
YES - each ingredient has a tare value assigned (the value is assigned on the basis of
data for a particular product saved to the products database);
•
NO - tare is not applied.
•
Means of ingredient verification:
Use this function to specify how to determine ingredient mass during formulation making
process, by TOLERANCE or by THRESHOLDS.
•
TOLERANCE: the software accepts mass of an ingredient as correct if it is within set
percent tolerance of total ingredient's mass (±%) - (data acquired from database of
products).
•
THRESHOLDS: the software accepts mass of an ingredient as correct if it is within set
thresholds (data acquired from database of products).
Specified method is in effect for all ingredients in the formula making process.
If during formulation making process one of the ingredients is weighed in excess (its
mass is out of high tolerance limit) the following message is displayed upon ingredient
weight
FORMULATION?>. Confirm, the software automatically recalculates mass of other
ingredients so that the mixture proportions are correct.
In order to make THRESHOLDS option active, the product/ingredient related data in
database must correspond with value selected for <MEANS OF INGREDIENT
VERIFICATION>
<THRESHOLDS> value is set for <MEANS OF INGREDIENT VERIFICATION>
parameter, otherwise particular ingredients mass recalculation is not possible, for
example when ingredient thresholds are specified and <TOLERANCE> value is set.
•
Multiplier's editing:
Function designed to enable making multiple mixtures according to a selected formulation in a
single weighing process:
•
YES - on selecting a formulation for processing the software requests for multiplier's
value, the provided value is in effect for all ingredients in the formulation making process,
i.e. each ingredient is multiplied by the value.
•
NO - multiplier value is disabled, default value is set to <1>.
•
Printout; for information on how to operate this function read section 15.7, Additional
Parameters for a Weighing Process.
22.2. Formulations – Quick Access Buttons
Each working mode features automatically displayed set of default on-screen quick access buttons.
The set can be modified, i.e. you can add out-of-list quick access buttons to it. Such operation
requires particular permissions level.
confirmation:
<VALUE
parameter,
OUT
OF
RANGE.
e.g.
ingredient
thresholds
- 91 -
RECALCULATE
THE
are
specified
and