3. Enter <500> value.
4. Confirm. Menu is displayed.
34.13.
Doorway Degree
Parameter for setting how wide the door is to be opened in case of an automatic operation.
Default value: <100 %> - maximally wide open door.
Procedure:
•
Enter <
Misc.> submenu.
•
Select <Doorway degree> parameter, an edit box is displayed.
•
Set <75%> value.
•
Confirm. The software returns to menu window.
34.14.
Automatic Ionization
Parameter for switching the ionization on and setting the ionization state,
options for selection: None/Low/High.
Procedure:
•
Enter <
Misc.> submenu.
•
Select <Automatic ionization> parameter, settings box is displayed.
•
Set <Low> value.
•
Confirm. The software returns to menu window.
34.15.
Autotest
<AUTOTEST> function has been designed to help you assess balance operation
and to diagnose the cause of weighing errors exceeding the maximum permissible
values.
AUTOTEST facilitates regular optimisation of balance settings which is necessary in
order to maintain the best possible repeatability and weighing time at the
workstation. The function allows you to monitor the parameters at optional moment
and save records from carried out tests.
There are two modules:
•
•
Prior each test the balance controls level status, temperature and humidity.
AUTOTEST FILTER;
AUTOTEST GLP.
- 172 -