Example:
Command:
UG<CR><LF>
Response:
UG_ct_OK<CR><LF>
Activate Sound Signal
Command overview:
Command activates BEEP sound signal for a specified amount of time.
Format: BP_TIME <CR><LF>
Response options:
BP_OK <CR><LF>
BP_E" <CR><LF>
BP_I <CR><LF>
TIME - parameter, decimal value specifying how long shall the sound last, parameter given in [ms].
Recommended range <50 ÷ 5000>.
If value greater than the permissible high limit is given, then BEEP sound is operated for the
maximum permissible amount of time.
Example:
Command:
BP_350<CR><LF>
Response:
BP_OK<CR><LF>
Caution:
BEEP sound activated via BP command is inhibited if in-course of its activation the sound gets
activated by means of other device: keypad, touch panel, proximity sensors.
Send All Implemented Commands
Format: PC CR LF
Command:
PC CR LF
Response:
PC_A_"Z,T,S,SI..."
Give Balance Type
Format: BN <CR><LF>
Response options:
BN_A_"x" <CR><LF>
BN_I <CR><LF>
x - balance series (inserted in between inverted commas), with general balance type in front.
Example:
Command:
BN <CR><LF>
Response:
BN_A_"AS"
Give Max Capacity
Format: FS <CR><LF>
Response options:
FS_A_"x" <CR><LF>
FS_I <CR><LF>
x - Max capacity value (in between inverted commas).
Example:
Command:
FS <CR><LF>
Response:
FS_A_"220.0000"
- command carried out, BEEP activated
- no parameter or incorrect format
- command understood but not accessible at this moment
- command understood, response: balance type
- command understood but not accessible at this moment
- command understood, response: Max capacity
- command understood but not accessible at this moment
- return current unit
- currently set unit is „ct"
BP_350<CR><LF> - activate BEEP for 350 ms
- BEEP activated
- send all implemented commands
- command carried out, all implemented commands
have been sent.
- give balance type
- balance type - "XA 4Y"
- return Max capacity
– Max capacity: "220 g".
- 193 -