AUTOTEST FILTER
Procedure consisting in loading and unloading of the internal weight for all possible
'filter' and 'value release' settings, the test requires 10 repetitions. It controls 2
parameters: 'Repeatability' and 'Stabilization time'.
The test takes approximately 1 hour. Upon completion, information regarding results
is obtained. Optimum settings for particular ambient conditions are provided.
Autotest Filter enables:
Upon completed autotest, a summary with results is displayed.
Information regarding filters settings is provided automatically via display of respective pictograms:
settings providing the fastest measurement (the shortest measurement time).
settings providing the most accurate measurement (the lowest possible value of
deviation for 10 measurements).
settings providing optimal measurement (the lowest value of time and deviation
product).
current settings for filters.
Measurement results:
*Filter type.
*<Value release> parameter value.
*Repeatability of indications expressed as standard deviation.
*Average time of measurement stabilization.
Example of Autotest Filter results:
You can quickly modify 'filter' and 'value release' settings. To do that, select results for particular
settings that are to be saved. Press <Activate>.
to obtain the highest measurement repeatability and the shortest
measurement time with acceptable value of repeatability; test results are
stored in balance memory until it is turned off,
test results printout,
direct and fast selection of optimal parameters.
- 173 -