HOW TO OPERATE PRINTOUTS
1. In order to print variables of GLP PRINTOUT section, press PRINT button on the operation panel,
only variables with YES attribute can be printed (see the above list).
2. Variables with YES attribute, present in the HEADER and FOOTER, are printed ONLY upon
pressing PRINT HEADER and PRINT FOOTER pictograms. The pictograms have to be added to
the bottom bar of the display as quick access buttons.
(For detailed instruction on how to add quick access buttons to the bottom bar of the display go to
section 14.3)
Print header data
Caution! Measuring units for mass indication:
Net – the main measuring unit (adjustment unit)
Tare – the main measuring unit (adjustment unit)
Gross – the main measuring unit (adjustment unit)
Current result – currently displayed measuring unit
Supplementary unit – supplementary measuring unit
Mass – the main measuring unit (adjustment unit)
Non-standard printout
Non-standard printout may contain TEXTS and VARIABLES (acquired from the software during
printout). Each non-standard printout is a separate project, featuring specific name by which it is
identified, and saved to the database of printouts.
Procedure:
1. Press <Non-standard printout> field.
2. Press <
Add> button, window featuring Name/Code/Project data opens.
3. Set name and code of a printout.
4. Press <Project> button.
5. Printout edit box with an on-screen keyboard is displayed.
6. Use the keyboard to design printout template, the printout may contain texts and variables.
Caution! It is possible to add a printout by import of configured texts from flash drives connected to
balance's USB port.
Printout's name is ONLY a NAME, and it is not part of its content.
For information on how to design a non-standard printout template read section 30.19 Printouts.

14.6. Profiles

Profiles function is described in section 13, Profiles.
15.

WEIGHING

Load the weighing pan. Wait for a stable indication (display of stability marker,
measurement result.
Record/printout of the measurement result is available on pressing <PRINT> key:
for verified balances – only stable measurement result is saved or printed (stability marker
visible on balance's display),
for non-verified balances – stable and unstable measurement results are saved or printed
(regardless of stability marker absence). For unstable measurement result question mark <?>
is printed next to the mass value.
Print footer data