Specifications

Synchronization Interfaces for FTB-8105/15/20/30

Synchronization Interfaces for
FTB-8105/15/20/30
SYNCHRONISATION INTERFACES
External Clock DS1/1.5M External Clock E1/2M
Tx Pulse Amplitude
2.4 to 3.6 V
Tx Pulse Mask
GR-499 figure 9.5
Typical power dBdsx
+0.6 dBdsx (0-133 ft)
Tx LBO
+1.2 dBdsx (133-266 ft)
Preamplification
+1.8 dBdsx (266-399 ft)
+2.4 dBdsx (399-533 ft)
+3.0 dBdsx (533-655 ft)
TERM: ≤ 6 dB (cable loss only)
Rx Level
Sensivity
(at 772 kHz for T1) DSX-MON:
≤ 26 dB (20 dB resistive loss
+ cable loss ≤ 6 dB)
Bridge: ≤ 6 dB (cable loss only)
Transmission Bit Rate 1.544 Mbit/s ± 4.6 ppm
Reception Bit Rate
1.544 Mbit/s ± 50 ppm
Intrinsic Jitter (Tx)
ANSI T1.403 section 6.3
GR-499 section 7.3
Input Jitter
AT&T PUB 62411
Tolerance
GR-499 SECTION 7.3
Line Coding
AMI and B8ZS
Input Impedance
75 ohms ± 5%,
(Resistive Termination) unbalanced
a
Connector Type
BNC
NOTES
a. Adaptation cable required for BANTAM.
b. SFP/XFP transceivers comply with IEC 60825 and 21 CFR 1040.10
(except for deviations pursuant to Laser Notice 50, dated July, 2001), for Class 1 or 1M lasers.
572
External Clock E1/2M
3.0 V
2.37 V
G.703 figure 15
G.703 figure 15
TERM: = ≤ 6 dB (cable loss only)
TERM: = ≤ 6 dB (cable loss only)
MON: ≤ 26 dB (20 dB resistive loss
MON: ≤ 26 dB (resistive loss
+ cable loss ≤ 6 dB)
+ cable loss ≤ 6 dB)
Bridge: ≤ 6 dB (cable loss only)
Bridge: ≤ 6 dB (cable loss only)
2.048 Mbit/s ± 4.6 ppm
2.048 Mbit/s ± 4.6 ppm
2.048 Mbit/s ± 50 ppm
2.048 Mbit/s ± 50 ppm
G.823
G.823
section 6.1
section 6.1
G.823 section 7.2
G.823 section 7.2
G.813
G.813
AMI and HDB3
AMI and HDB3
75 ohms ± 5%,
75 ohms ± 5%,
unbalanced
unbalanced
a
BNC
BNC
ETHERNET ADD/DROP INTERFACE
Trigger 2 MHz
10/100/1000 Base-T (Add/Drop)
0.75 to 1.5 V
Compliance
G.703 figure 20
Connector
Gigabit Ethernet (Add/Drop)
Interface/connector
Compliance
Wavelength/Max Tx level
≤ 6 dB (cable loss only)
REF-OUT INTERFACE
Parameter
Tx pulse amplitude
Transmission frequency
G.703
table 11
Clock divider = 16
Clock divider = 32
Clock divider = 64
Output configuration
75 ohms ± 5%,
Load impedance
unbalanced
Maximum cable length 3 meters
BNC
Connector Type
10 Mbit/s: IEEE 802.3 section 14
100 Mbit/s: IEEE 802.3 section 25
1000 Mbit/s: IEEE 802.3 section 40
RJ-45 Ethernet
SFP/Dual LC
1000 Mbit/s: IEEE 802.3 Section 40
850, 1310 nm/—3 dBm
1550 nm/+5 dBm
Value
600 ± 150 mVpp
SONET/SDH/ 10 GigE LAN OTU2
OTU1e
10 GigE WAN
622.08 MHz
644.53 MHz 669.33 MHz 690.57 MHz 693.48 MHz
311.04 MHz
322.266 MHz 334.66 MHz 345.29 MHz 346.74 MHz
155.52 MHz
161.133 MHz 167.33 MHz 172.64 MHz 173.37 MHz
AC coupled
50 ohms