EXFO FTB-8510B Guia Del Usuario página 926

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

DSn Tabs
DS1/1.5M TX
Error Injection
Allows manual or automated error injection.
Note: Error injection is not available when the framing is set to Unframed.
Type: The following error types are available with both manual and
automated injection modes. Choices are Framing Bit and CRC-6.
CRC-6 is only available with ESF framing.
Note: Choices depend on the selected framing.
Amount: Select the amount of error to be generated. Choices are 1
through 50. The default setting is 1.
Send button: Press Send to manually generate error(s) according to
the Error Type and the Amount of Errors selected.
Rate: Press Rate to select the injection rate for the selected error. The
rate must be within the minimum and maximum values specified.
Continuous: Generates the selected error to its theoretical maximum
when the Continuous check box is selected. The Continuous check
box is cleared by default.
On/Off button: The On/Off button is used to activate/deactivate the
selected automated error at the rate specified or at its theoretical
maximum rate when the Continuous check box is selected. This
setting is disabled (Off) by default.
276
FTB-8100 Series Transport Blazer

Publicidad

Capítulos

Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

Ftb-8510gFtb-8525Ftb-8535Ftb-8120nceFtb-8103nce

Tabla de contenido