OTN Tabs

ODU TX

ODU TX
Press TEST, ODUk, and ODU (under ODUk TX).
Error Injection
Allows manual or automated error injection.
Type: The following error types are available with both manual and
automated injection mode: ODU-BIP-8, and ODU-BEI. The default
setting is ODU-BIP-8.
Amount: Select the amount of errors to be generated. Choices are 1
through 50. The default setting is 1.
Send button: Press Send to manually generate error(s) according to
the Error Type and the Amount of Errors selected.
Rate: Press Rate to select the injection rate for the selected error. The
rate must be within the minimum and maximum values specified. The
default setting is 6.5E-05.
Continuous: Generates the selected error to its theoretical maximum
rate when the Continuous check box is selected. The Continuous
check box is cleared by default.
On/Off button: The On/Off button is used to activate/deactivate the
selected automated error at the rate specified or at its theoretical
maximum rate when the Continuous check box is selected. This
setting is disabled (Off) by default.
180
FTB-8100 Series Transport Blazer