Creating and Starting a Test Case
Typical Test Cases
2. Interface connector selection:
2a. From the data path
2b. For Decoupled test
2c. Press Next.
3. Interface selection and configuration:
3a. Press the desired interface: STS-3e, STS-1e, STM-1e, or STM-0e.
3b. Select the Line Coding, TX LBO, and RX
3c. Press Next.
80
selector, press the
BNC electrical interface connector.
mode, both TX and
RX ports have to be
selected and
configured:
First select the
interface type for
TX from the data
path selector then
proceed with the
rest of the test setup steps to set the test parameters for the TX
interface. At the end, do not press Finish yet. Press Back to return
to the RX/TX selection screen and select the interface type for RX
from the data path selector then proceed with the rest of the test
setup steps to set the test parameters for the RX interface.
Termination Mode (Term or Mon). For more
information, refer Port TX (Electrical Interfaces)
on page 138 for Line Coding and LBO, to RX -
DSn Tabs on page 269 or PDH Tabs on page 369
for Termination Mode.
FTB-8100 Series Transport Blazer