Radiology

1. MRI systems that meet the following criteria:
• MRI magnet strength of 1.5T only, in a horizontal closed bore system (no vertical-field,
standing, or extremity systems).
• Gradient systems with a maximum gradient slew rate per axis less than or equal to 200 T/m/s.
• Maximum spatial field gradient less than or equal to 40 T/m (4000 gauss/cm).
2. MRI coil setup:
• Transmit coil: 1.5T Full Body transmit/receive, Head transmit/receive, or Extremity transmit/
receive. RF quadrature
• Receive-only coil: Any type.
• Hydrogen/proton imaging only.
3. Patient status and positioning:
• The patient is in supine or prone position only.
• The Lead implant location is epidural.
• The IPG is implanted in the upper buttock or the lower flank.
• The patient has enabled MRI mode using the Remote Control.
1 RF Quadrature coil – RF Quadrature Coils produce an RF field with circular polarization perpendicular to the static magnetic field.
ImageReady™ MRI Full Body Guidelines for WaveWriter Alpha™ and WaveWriter Alpha™ Prime Spinal Cord Stimulator Systems
only.
1
MR Conditions of Use
92395577-02
7 of 797