MR System Preparation
NOTE: The Stimulator will retain the stimulation and program settings that were set before MRI Mode was enabled. If the
stimulation was ON before MRI Mode was enabled, then disabling MRI Mode turns the stimulation back ON. If stimulation
was OFF before MRI Mode was enabled, then disabling MRI Mode keeps the stimulation OFF.
9. The Home Screen on the Remote Control will not display the MR Conditional Symbol once MRI
Mode is disabled.
ImageReady™ MRI Full Body Guidelines for WaveWriter Alpha™ and WaveWriter Alpha™ Prime Spinal Cord Stimulator Systems
92395577-02
24 of 797