MR System Preparation

MR System Preparation
Table 3 summarizes the MR Scanner-related Conditions of Use that must be met in order for an MR
Conditional scan to be performed. For each condition or requirement, recommended actions to determine
conformance are listed.
Table 3. MR System Conditions
#
Condition for Scanning
1.
MRI systems that meet the following criteria:
• MRI magnet strength of 1.5T only, in a horizontal closed
bore system (no vertical-field, standing, or extremity
systems).
• Gradient systems with a maximum gradient slew rate per
axis less than or equal to 200 T/m/s.
• Maximum spatial field gradient less than or equal to
40 T/m (4000 gauss/cm).
2.
MRI coil setup:
• 1.5T Full Body transmit/receive coil, head transmit/receive
coil, or Extremity transmit/receive coil
• Receive only coil: Any type
• Hydrogen/proton imaging only
ImageReady™ MRI Full Body Guidelines for WaveWriter Alpha™ and WaveWriter Alpha™ Prime Spinal Cord Stimulator Systems
92395577-02
16 of 797
Actions
Check the technical specifications of the MRI Scanner.
Check the technical specifications of the MRI Coil.