To reduce the risk of cross-contamination
• Ensure the disposable polyethylene barrier protection sleeve has been correctly
applied over the entirety of the attachment and handpiece before beginning
CAUTION
a procedure.
• The polyethylene barrier sleeve does not replace cleaning and disinfection of
the dental instruments. Please clean and reprocess dental instruments after each
patient as described in Section 4, Hygiene and Disposal.
• Do not apply the SmartLite Pro EndoActivator™ barrier sleeve too tightly over
the attachment head to avoid motor stalling or obstructing tip oscillation. Briefly
check proper tip operation outside the patient`s mouth before intraoral use.
3.3 Operation – Activator Tip Selection
1. Prepare canal to produce a fully tapered shape.
2. Select the activator tip that manually fits loosely within 2mm of working length.
See size chart below.
To reduce risk of tissue damage due to extrusion of irrigant
• Only operate the tip 2mm or more from working length or apex.
• Small and Medium (22mm) tips are marked with depth gauge rings at 18, 19
and 20mm.
CAUTION
• Medium Long (28mm) tip is marked with depth gauge rings at 18, 19, 20, 24, 25
and 26mm.
• Avoid use when apex is incomplete or open.
Tip Size
Small
Medium
Medium Long
3. With clean/disinfected gloves, select a new, unused activator tip. Ensure tip is in the original,
undamaged packaging. If packaging is damaged or unsealed, do not use the contaminated
activator tip.
To reduce the risk of cross-contamination
• Ensure the activator tip is in the original, undamaged packaging.
• If packaging is damaged, dispose of activator tip and damaged packaging
CAUTION
according to local regulations.
• Dispensing materials with clean/disinfected gloves in a separate room, bringing
only what will be used into the operatory is strongly recommended.
• Prior to use, process tips as described in Section 4, Hygiene and Disposal.
4. Completely peel off the seal from the blister back. Use sterile college pliers or adequate
forceps to grab the activator tip at the colored shaft and remove from its packaging away
from the patient field. Prior to use, follow the Processing Instructions for activator tips.
5. While holding the handpiece in one hand and stabilizing the head of the attachment with
the thumb, attach the activator tip over the barrier sleeve in place protecting the assembled
attachment and handpiece. The activator should snap on firmly, promoting a secure
connection with the attachment.
Small parts - To reduce the risk of inhalation or aspiration
• Ensure activator tip is firmly attached.
• Do not use with lateral pressure.
CAUTION
• Do not use with instruments other than designated activator tips.
• Rubber dam isolation is strongly recommended.
6 Adjust the EndoActivator Attachment: The attachment is rotatable by 360°, thus the
position of activator tip may be individually adjusted.
12
Length
22mm
22mm
28mm
Tip Diameter
0.15mm
0.25mm
0.25mm
Taper
15.02
25.04
25.04