4.2 Handpiece, Attachment, Charging Base
To reduce the risk of Electrical short-circuit or dangerous malfunction
Injury
•Safeguard handpiece against liquid penetration during cleaning and disinfection.
CAUTION
NOTICE
Wrong cleaning or disinfection method
To reduce the risk of damage to components
• Follow Instructions Below.
Instructions for Cleaning and Disinfecting Handpiece, Attachment, Charging Base
Warnings
• The handpiece, attachment and charging base are not sterilizable by autoclave.
• The system components cannot tolerate high-level disinfection procedures.
Intermediate-level disinfection is appropriate for the handpiece, attachment and
charging base.
• Do not autoclave in steam autoclave.
• Do not clean/disinfect in automated washer/disinfector.
• Do not immerse in liquid.
• Do not clean or disinfect with chlorine bleach/sodium hypochlorite (corrosion of
contacts) or Lysol® Brand I.C.™ Disinfectant Spray (cracking of charging base).
• Disconnect the power supply plug from the power outlet and charging base unit prior
to cleaning/disinfection.
Limitations on
• Repeated reprocessing has minimum e ect on these instruments. End of life is normally
reprocessing
determined by wear and damage due to use.
• Cold liquid immersion disinfection/sterilization, chemical vapor sterilization and dry
heat sterilization methods have not been tested or validated for e cacy and are not
recommended for use.
Initial treatment
• Remove barrier sleeve and discard according to local regulations.
at the point
• Use a new, clean pair of examination gloves.
of use
• Do not disassemble attachment from handpiece at point of use.
• Wipe vigorously with disposable cloth / paper wipe in combination with an alcohol-
based, tuberculocidal, quaternary ammonium solution with a label claim for cleaning.
• Remove all visible soil, ensuring fluid penetrates all crevices. Use fresh wipes to rub fluid
into the crevices. Do not allow solution to penetrate the casing. Discard used wipes.
Additional wipes may be used.
• Do not remove battery pack from handpiece. Do not attempt disassembly of
charging base.
• It is recommended that the device be reprocessed as soon as is reasonably practical
following use.
• Start reprocessing within 1 hour after use.
• Charging base should be reprocessed as soon as reasonably practical after being
exposed to spatter or spray of body fluids or touched by contaminated hands or
contaminated handpiece.
Preparation
Always disassemble attachment from handpiece before processing.
before cleaning
Use moderate force to pull the attachment from the handpiece.
Cleaning and
Do not use automated washer/disinfectors for reprocessing system components.
Disinfection:
Component damage will occur.
Automated
Cleaning:
Cleaning by Immersion
Mechanical
N/A – No validated processes. Do not subject components to immersion. Component
damage will occur.
15