Guidance and manufacturer's declaration – electromagnetic immunity
The SmartLite Pro® System is intended for use in the electromagnetic environment specified below. The customer or the
user of the SmartLite Pro® System should assure that it is used in such an environment.
IMMUNITY Test
IEC 60601
Test level
Electrostatic
± 8 kV contact
discharge (ESD)
± 2, ± 4, ± 8, ± 15 kV air
IEC 61000-4-2
Electrical fast
± 2 kV for power supply
transient/burst
lines
IEC 61000-4-4
± 1 kV for input/output
lines
Surge
± 0.5, + 1 kV line(s) to
line(s)
IEC 61000-4-5
± 0.5, ± 1, ± 2 kV Line(s)
to earth
0% U
Voltage dips, short
interruptions and
voltage variations
0% U
on power supply
input lines
70% U
IEC 61000-4-11
0% U
cycles
Power frequency
30 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
Compliance level
± 8 kV contact
± 2, ± 4, ± 8, ± 15 kV air
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 0.5, + 1 kV line(s) to
line(s)
± 0.5, ± 1, ± 2 kV Line(s)
to earth
0% U
for 0,5 cycle
T
for 1 cycle
0% U
T
for 25/30 cycles
70% U
T
for 250/300
0% U
T
cycles
30 A/m
Electromagnetic
environment - guidance
Floors should be wood, concrete
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
for 0,5 cycle
Mains power quality should be
T
that of a typical commercial or
hospital environment. If the user
for 1 cycle
T
of the SmartLite Pro® System
requires continued operation during
for 25/30 cycles
power mains interruptions, it is
T
recommended that the SmartLite
for 250/300
Pro® System be powered from an
T
uninterruptible power supply or
a battery.
Power frequency magnetic fields
should be at levels characteristic
of a typical commercial or
hospital environment.
21